16:30 |
139. |
Z-Selective Multi-Spectral 3D Imaging: A MAVRIC-SEMAC Hybrid
Kevin M. Koch1,
Kevin F. King1, Brian A. Hargreaves2,
Graeme C. McKinnon1
1Applied Science Laboratory, GE
Healthcare, Waukesha, WI, United States; 2Department
of Radiology, Stanford University, Palo Alto, CA, United
States
Both the MAVRIC and SEMAC
methods have been shown capable of significantly reducing
susceptibility artifacts near metallic implants. Here, we
demonstrate that advantageous features of both methods can
be utilized in combination. In particular, the
z-selectivity of the SEMAC can be interfaced with the
encoding mechanisms and spectral overlap utilized by MAVRIC.
In doing so, a technique that offers increased volume
selectivity while maintaining smooth spectral image
combinations is demonstrated. Demonstrations of this hybrid
approach on phantom and in-vivo implant scenarios are
presented. |
|
|
|
16:42 |
140. |
Dipole-Based Filtering for Improved Removal of Background
Field Effects from 3D Phase Data
Samuel James Wharton1, Richard Bowtell1
1Sir Peter Mansfield Magnetic
Resonance Centre, University of Nottingham, Nottingham,
United Kingdom
A robust method for filtering
3D phase data dominated by rapidly spatially varying
externally generated fields is presented. One or more dipole
point sources situated outside of the region of interest are
used to model and remove the unwanted background fields
caused by remote tissue/air interfaces such as those that
are present in the sinuses. The method was tested on
simulated and experimentally acquired phase data and
compared to other commonly used filtering methods, including
Fourier filtering and polynomial fitting. The results show
that the dipole-based filter outperformed the other methods
in removing unwanted fields and preserving image contrast. |
|
|
|
16:54 |
141. |
Improved
Background Field Correction Using Effective Dipole Fitting
Tian
Liu1, Ildar Khalidov2, Ludovic de
Rochefort3, Pascal Spincemaille2, Jing
Liu2, Yi Wang1
1Biomedical
Engineering, Cornell University, New York, NY, United
States; 2Radiology, Weill Cornell Medical
College, New York, NY, United States; 3MIRCen,
I2BM, DSV, CEA, Fontenay-aux-Roses, France
Effective dipole fitting
removes susceptibility induced global background field. It
assumes that each independent voxel outside a given region
of interest (ROI) are responsible for the background field
inside that ROI. It removes the field generated by these
sources, while preserving the field arising from local
susceptibility variations inside the ROI. |
|
|
|
17:06 |
142. |
A Novel
Approach for Separation of Background Phase in SWI Phase
Data Utilizing the Harmonic Function Mean Value Property
Ferdinand Schweser1, Berengar Wendel Lehr2,
Andreas Deistung2, Jürgen Rainer Reichenbach2
1Medical Physics Group,
Department of Diagnostic and Interventional Radiology, Jena
University Hospital, Jena, Germany; 2Medical
Physics Group, Department of Diagnostic and Interventional
Radiology, Jena University Hospital, Jena, Germany
In this contribution, we
present, for the first time, a non-heuristic, parameter-free
approach for high-precision separation of local phase and
background phase contributions for in vivo SWI-data. |
|
|
|
17:18 |
143. |
Analysis
of Quadratic Field Distortions Using the Fractional Fourier
Transform
Carlos Sing-Long1,2, Vicente Parot1,2,
Carlos Lizama3, Sergio Uribe, 2,4,
Cristian Tejos1,2, Pablo Irarrazaval1,2
1Department of Electrical
Engineering, Pontificia Universidad Catolica de Chile,
Santiago, Chile; 2Biomedical Imaging Center,
Pontificia Universidad Catolica de Chile, Santiago, Chile;
3Department of Mathematics and Computer Science,
Universidad de Santiago de Chile, Santiago, Chile; 4Department
of Radiology, Pontificia Universidad Catolica de Chile,
Santiago, Chile
In Magnetic Resonance Imaging
(MRI) the distortions produced by field inhomogeneities can
be corrected with post processing techniques, e.g. linear
correction and conjugate phase reconstruction methods.
However, these methods do not provide a theoretical
framework to analyze the distortions. In this work, we
propose the Fractional Fourier Transform (FrFT) as a way to
study the distortions produced by quadratic field
inhomogeneities. We analyze some commonly used sequences to
exemplify the usefulness of this method. We also show how
this analysis can be used to reconstruct artifact-free
images obtained from non homogeneous fields. |
|
|
|
17:30 |
144. |
Generalized Non-Linear SENSE Shimming
Daniel Nicolas Splitthoff1,
Maxim Zaitsev1
1Dept.
of Diagnostic Radiology, Medical Physics, University
Hospital Freiburg, Freiburg, Germany
With the SENSE Shimming (SSH)
approach a method was introduced recently that allows for
estimating B0 field inhomogeneities based on a reference
image and a series of points on a single free induction
decay (FID). In the original approach the temporal evolution
of the FID data is explained by field inhomogeneities, using
linear approximations. Effects caused by relaxation and
those caused by inhomogeneities can therefore not be
distinguished and values can only be given relative to a
baseline measurement. We here present an extension to the
method, which takes into account a larger range of the FID
in order to explain not only B0 inhomogeneities but
relaxation as well and which therefore allows for accurate
field map estimation based on a reference image and a single
FID. Since the signal equation is non-linear, the linear
fitting of the original approach has to be replaced by a
non-linear optimization. The feasibility of the method is
shown on in vivo data. |
|
|
|
17:42 |
145. |
On the
Feasibility of Single-Shot EPI During Higher-Order Shim
Settling
Signe
Johanna Vannesjö1, Lars Kasper1,
Matteo Pavan1, Christoph Barmet1,
Klaas Paul Pruessmann1
1Institute for Biomedical
Engineering, ETH and University Zürich, Zürich, Switzerland
Susceptibility artefacts is a
major problem in MRI, becoming more severe with higher field
strengths and longer read-out trajectories. Updating the
shim settings between acquisition of different slices allows
for optimizing the shims to smaller subvolumes, but puts
high requirements on the timing characteristics of the shim
switching. Here the settling dynamics of the higher order
shims were measured using a 3rd order dynamic field camera.
Long-living (seconds) eddy currents were found, which had a
significant effect on image quality. Based on measured
k-space trajectories, it was possible to reconstruct phantom
images acquired during eddy current settling. |
|
|
|
17:54 |
146. |
Increasing Spoiling Efficiency in RF-Spoiled Gradient Echo
Sequences by Averaging of Phase-Cycle Adapted K-Spaces
- not available
Jochen Leupold1, Jürgen Hennig1
1Dept. of Diagnostic Radiology,
Medical Physics, University Hospital Freiburg, Freiburg,
Germany
RF-spoiled gradient echo
sequences (FLASH, SPGR, T1-FFE) require a spoiler gradient
in order to suppress ghost artefacts. Here we show that two
k-spaces can be adapted to the RF phase cycle such that
averaging of them leads to elimination of these artefacts
even if the spoiler gradient has only half of the moment
that is requried for common RF-spoiled gradient echo
acquisition. |
|
|
|
18:06 |
147. |
Transient
RF Spoiling for 3D Look-Locker Acquisitions
Trevor Wade1,2, Charles McKenzie1,3,
Brian Rutt4
1Imaging Research Laboratories,
Robarts Research Institute, London, ON, Canada; 2Biomedical
Engineering, The University of Western Ontario, London, ON,
Canada; 3Medical Biophysics, The University of
Western Ontario, London, ON, Canada; 4Diagnostic
Radiology and Richard M Lucas Center for Imaging, Stanford
University, Stanford, CA, United States
Theoretical and experimental
investigation of RF spoiling in the special case of
accelerated 3D Look-Locker imaging has led to an improved
value for the phase increment used in the standard RF
spoiling scheme. Poor choice of phase increment leads to an
inversion recovery curve that deviates significantly from
the theoretical ideal, leading to an inaccurate estimate of
the recovery time constant. Simulations were used to
determine improved values for the phase increment based on
minimizing summed squared differences, or time constant
measurement accuracy. These were tested experimentally and
found to be superior to previously reported values for most
imaging parameters. |
|
|
|
18:18 |
148. |
Parallel
Imaging for Efficient Spike Noise Detection and Correction
Feng
Huang1, Wei Lin1, Yu Li1,
Arne Reykowski1
1Invivo Corporation,
Gainesville, FL, United States
Spike noise is a term used to
describe broadband electrical interference in an MRI
system. The result of spike noise can be seen in k-space as
a bright dot, which will translate into some type of
striping in the final image. Usually, the scan has to be
repeated if random spike occurs. A parallel imaging based
method, COnvolution and Combination OperAtion (COCOA), has
been proposed for non-rigid motion compensation. In this
work, it is shown that COCOA can be used to robustly detect
and correct random spikes in an efficient way. Hence
repeated scan can be avoided. |
|
|
|
|